Yonsei Medical Journal ; : 634-641, 2012.
Article in English | WPRIM | ID: wpr-22417

ABSTRACT

PURPOSE: To evaluate reliability of Fourier-domain optical coherence tomography (OCT) (RTVue), time-domain OCT (Visante), and slit-lamp adapted ultrasonic pachymetry (SL-US) in the measurement of central corneal thickness (CCT). MATERIALS AND METHODS: Thirty healthy volunteers visited our clinic 3 times and fifty eyes were measured by one physician. RTVue and Visante were randomly performed, and then SL-US, in which the ultrasound probe was inserted into the Goldmann tonometry mount, was done. During the second visit, each measurement was repeated 3 times. Measurements on the second visit were averaged, and agreement among the instruments was investigated with Bland-Altman plots. RESULTS: RTVue showed smaller repeatability coefficient than Visante and SL-US (4.7, 8.3, and 7.7 microm, respectively). Intersession reproducibility of RTVue and Visante was worse than their repeatability. CCT of RTVue was estimated to be maximally different by 11.8 microm from CCT of Visante and 8.8 microm from CCT of SL-US. The repeatability coefficient of SL-US was 7.7 microm and its reproducibility was similar to the repeatability. CONCLUSION: CCT measured by RTVue showed good reliability and generally agreed with Visante and SL-US. SL-US was as reliable as triple touching conventional US even with a single touch on the cornea.


Subject(s)
Adult , Female , Humans , Male , Cornea/anatomy & histology , Corneal Topography , Reproducibility of Results , Tomography, Optical Coherence/methods
SELECTION OF CITATIONS
SEARCH DETAIL